Inspection of our rigid endoscope

Before each use, conduct a thorough visual inspection and functional check of the endoscope:


1.Ensure that the endoscope is in perfect condition before use. Only use endoscopes without any signs of damage or wear. 


2.To check if the endoscope is functioning properly, examine the reflected light on the surfaces of the ocular and objective lenses. 


3.Inspect the quality of the fiber optics by holding the light post toward a light source and observing the distal tip. If the light is evenly distributed, the fiber optics are in good condition. Darker areas could indicate broken fibers. 


4.Foggy images may suggest that moisture has penetrated the seal, while partially or completely obstructed views typically indicate a damaged lens. 


By consistently performing these inspections and checks, you can ensure the endoscope's optimal performance and maintain the safety of both patients and users.